HOME       ABOUT US       PRODUCTS       INDUSTRY SOLUTION        PRINCIPALS       SUPPORT       ENQUIRY       CONTACT US  
 
UV-VIS - DUAL BEAM - UVD 2950
 
Spectro UV-Vis Double PC (Model UVD-2950) is a high performance UV-Vis double beam automatic scanning spectrophotometer. It is a two (2) cell spectrophotometer, now with a new and improved bandwidth of 1nm! . Model UVD-2950 spectrophotometer offers high performance, ease of use and reliability, which can be used in various applications. Spectrophotometer Model UVD-2950 can be used extensively for qualitative and quantitative analysis in such fields as pharmaceutical inspection, clinical analysis, petro-chemistry laboratories, chemistry and biochemistry laboratories, DNA/RNA analysis as well as in quality control departments, i.e. environmental control, water management, food processing, and agriculture.
 
Spectro UV-Vis Double PC (Model UVD-2950) utilizes a new optical system design and is microcomputer controlled. With its focused-beam design, the system provides optimal and reproducible results for small samples. The sample beam and the reference beam are provided within the same sampling space, facilitating a wider and longer scan of data, providing a more detailed view of the results in an easy to use environment. This instrument has excellent baseline stability and high resolution and permits scanning, quantitative analysis, kinetic spectrophotometric analysis and DNA/RNA analysis through PC control. This product is capable of processing data, from analytical and spectrum testing.

Spectro UV-Vis Double PC (Model UVD-2950) has a large LCD screen which displays the menu screen and of course makes the device user friendly. Additionally, this instrument permits the apparatus to be linked to a computer and a printer to display the photometric and spectral data on the PC monitor. This connection is controlled by the RS232 AND USB interface, compatible with Windows XP, & and 8, using the new UVWin 6.0 UV-VIS application software.

Labomed, Inc. is certified byISO 9001-2013, has CE Conformity and is FDA Licensed. Spectro UV-Vis Double PC (Model UVD-2950) with fixed bandwidth of 1 nm is a high-performance, reliable, and exceptional value instrument which is the hallmark of Labomed UV-Vis spectrophotometers.
 
Features:

Baseline Stability: The Double beam monitoring ratio system enhances baseline stability.

1nm fixed bandwidth New!: The Double beam monitoring ratio system enhances baseline stability.

Excellent Resolution: The big-caliber light path enhances the instrument’s energy, reduces its noise and raises its resolution performance.

2 Cell Holder: Spectro UVD-2950 has 2 cell holders for reference (standard) and sample.

User-friendly light source: The light source performs an automatic interchange, selectable within the working range of the light source. The socket deuterium lamps and tungsten lamps facilitate light source replacement, simplify maintenance and reduce operation error.

Convenient Display: The large backlit LCD screen displays both photometric values and spectral curves.

Full use of Computer Technology: Being computer controlled with RS-232 interface and working on the Windows platform with the UV/Win application software.

The key components are all adopted from the world famous manufacturer, such as deuterium lamp, silicon photodiode and holographic grating, which ensures the stabilization and credibility of the Instrument for extended life.

Computer System is optional (NOT INCLUDED).
 
Technical Specifications:
 
1) Wavelength Range: 190 nm – 1100 nm
- Spectral Bandwidth: 1.0 nm. New and Improved!
- Resolution: 1nm.
- Wavelength Display 0.1nm resolution
- Stray Light: >2.10Abs (200nm)
- Wavelength Accuracy: +0.3nm (with automatic wavelength correction)
- Wavelength Reproducibility: +0.2 nm
2) Photometric System: Double beam optical system
- Photometric Method: Transmittance, absorbance, energy and concentration
- Photometric Range: -0.3~3.0 Abs
±0.002Abs (0~0.5A)
- Photometric Accuracy: ±0.004Abs (0.5~1.0A)
±0.3%T (0~100%T)
±0.001Abs (0~0.5A)
- Photometric Reproducibility: ±0.002Abs (0.5~1.0A)
±0.15%T (0~100%T)
- Photometric Display: -9999 ---- 9999
- Photometric Noise: ±0.001Abs (at 500nm, 2nm Spectral Bandwidth 0 Abs)
- Scanning Speed: 1400nm/min
- Baseline Flatness: ±0.002 Abs
- Baseline Stability: 0.0008Abs/h (2 hours warming up, 2nm Bandwidth, 500nm)
- Slew Rate of Wavelength: 3600nm/min
3) DNA/RNA Measurement:  
- Results Printout: Printing of measured data by using any Printer with Parallel Port connection available.
4) Mainframe: Compact and standalone spectrophotometer mainframe
- Light Source: Socket Deuterium Lamp and Socket Tungsten Halogen Lamp.
- Detector: Double Beam
- Sample Chamber: 2 cell holder
- Display: Liquid Crystal Display (LCD 320 iÀ240 dot matrix)
- Keypad: Touch soft keys.
- PC Interface: PC Interface: RS-232
- Size: 22"x16"x10”
- Weight: 55 Lb
 
 
  VIDEOS
 
 
© LCGC Instruments